IVDR regulation: what non-EU IVD makers must do

Share this article

WhatsAppLinkedIn

Published on

The EU IVDR regulation, Regulation (EU) 2017/746, makes you as a manufacturer outside the EU responsible for your test: you classify it, involve a notified body where required, and appoint one authorised representative in the Union. At MAAK Advocaten in Amsterdam we review that role split and your contracts before a Dutch distributor or authority asks.

In short

  • As a manufacturer outside the EU you classify the test, involve a notified body unless it is non-sterile class A, and appoint one authorised representative (Articles 11, 47, 48).
  • Your old IVDD status carries on only until 31 December 2027, 2028 or 2029, by route and class, and only if the conditions of Article 110(3c) are met.
  • In the Netherlands the inspectorate (IGJ) supervises, and instructions must be in Dutch, with a narrow exception for professional-only devices.
  • Have a Dutch product law lawyer review your mandate and any own-brand agreement before you sign.

This page is part of our overview of sector-specific product rules in the EU and of EU product law.

Does the IVDR regulation apply to your test?

It applies to a product you intend for the in vitro examination of specimens from the human body, solely or principally to give information on a condition, a predisposition or a treatment (Regulation (EU) 2017/746, Article 2, point 2).

Your intended purpose follows from your label, instructions and promotion (Article 2, point 12). A device that incorporates a medical device as an integral part falls under the MDR, and the IVDR requirements apply to its in vitro diagnostic part (Article 1(4)); see our page on the EU MDR in the Netherlands.

It applies since 26 May 2022 (Article 113(2)) and replaced Directive 98/79/EC (IVDD), except what Articles 110 and 112 keep running. A Commission proposal to amend it, COM(2025) 1023, is not law (see the opinion of the Economic and Social Committee).

Which class does your test fall into, and do you need a notified body?

The IVDR sorts devices into classes A, B, C and D by intended purpose and inherent risk (Article 47(1), Annex VIII). Only non-sterile class A is self-declared; classes B, C and D involve a notified body (Article 48).

ClassAnnex VIII, in shortRoute
Dtransmissible agents in blood or tissue for transfusion; ABO, Rhesus, Kell, Kidd, Duffynotified body
Csexually transmitted agents, companion diagnostics, cancer, genetic testsnotified body
Bresidual category; some self-testsnotified body
Ageneral laboratory use, instrumentsown declaration; if sterile, limited notified body

Classifying and building the file are technical work for your regulatory team and notified body; we do not assess the class of an individual test. See our page on the notified body and conformity assessment; you may not apply to a second notified body for the same procedure (Article 49(1)).

Who must do what, and who is your authorised representative?

You carry the manufacturer’s duties of Article 10 yourself. If you are not established in a Member State, you may place a device on the EU market only after appointing one authorised representative with a written mandate (Article 11(1) and (2)).

RoleMain dutyArticle
Manufacturerdesign, risk management, technical file, declaration, CE, UDI10
Authorised representativeone per manufacturer; written mandate; checks declaration and file11, 12
Importerplaces only compliant devices; checks CE, declaration, representative13
Distributordue care; checks CE, declaration, information, UDI14
Own brand, changed purpose or deviceimporter or distributor becomes manufacturer16

The mandate cannot hand your core duties to the representative (Article 11(4)). Your authorised representative, who must be established in the Union (see our page on the EU authorised representative in the Netherlands), is jointly and severally liable with you for defective devices if you fail Article 10 (Article 11(5)).

What manufacturers outside the EU most often get wrong is thinking their authorised representative takes the compliance work over. Technical documentation, declaration, UDI, quality system and corrective action stay with you.

Note: A consultant offers you a standard mandate to act as your authorised representative. Article 11(5) makes the representative jointly and severally liable with you if you fail Article 10, whatever the mandate says; the mandate decides recourse between you. Have a Dutch product law lawyer review the mandate, including recourse, before you sign it.

Does your IVDD certificate or self-declaration still carry you?

Only under the conditions of Article 110: a device with a valid IVDD certificate until 31 December 2027, a self-declared device that now needs a notified body until 2027, 2028 or 2029, by class.

RouteOn the market untilApplication byWritten agreement by
IVDD certificate (Article 110(3a))31 December 202726 May 202526 September 2025
Self-declared, now class D (Article 110(3b))31 December 202726 May 202526 September 2025
Self-declared, now class C (Article 110(3b))31 December 202826 May 202626 September 2026
Self-declared, now class B or class A sterile (Article 110(3b))31 December 202926 May 202726 September 2027

Article 110(3c) adds conditions: no significant change to design or intended purpose, no unacceptable risk, and a quality system under Article 10(8) by 26 May 2025. For class D and class C these dates have passed.

A second point manufacturers underestimate is that their old IVDD status simply carries on. It depends on conditions and dates: a quality system, an application to a notified body and a written agreement.

Note: Your Dutch distributor asks whether your test may still be sold after 2027. Whether Article 110(3c) is met is a legal reading of dates and documents. Have a Dutch product law lawyer check your evidence before you sign.

Is a Dutch distributor asking about your IVDR file? Put the file and the contract to our product law partner in the Netherlands.

What must you register, mark and supply for the Dutch market?

Draw up the EU declaration of conformity, affix CE, carry a UDI, register as an operator and supply the information in Dutch. These duties are yours; your importer checks them (Article 13).

Instructions and labelling must be in Dutch; English is allowed only for a device used exclusively by healthcare professionals with an adequate command of English. Documents for an authority may be in Dutch or English (Regeling medische hulpmiddelen, Article 1). A self-test is meant for lay persons (Article 2, point 5), so that exception does not apply automatically.

Commission Decision (EU) 2025/2371 of 26 November 2025 confirms four Eudamed systems as functional; obligations tied to a system apply six months after that notice (Article 113(3)(f)). If you expect to stop supplying a device and that could seriously harm patients or public health in one or more Member States, you inform the authority of the Member State where you or your representative are established, and the customers you supply directly, at least six months ahead, other than in exceptional circumstances (Article 10a).

Who supervises the IVDR in the Netherlands, and what if the inspectorate writes?

The inspectors of the Health and Youth Care Inspectorate (IGJ) supervise compliance with the Wet medische hulpmiddelen (Dutch Medical Devices Act) and the IVDR. The Minister can impose an order subject to a penalty and an administrative fine (Wet medische hulpmiddelen, Articles 10, 12 and 14).

You or your representative must keep the documentation available to the Minister even after bankruptcy or cessation (Wet medische hulpmiddelen, Article 5b(1)). A decision is challenged by objection and appeal within six weeks (Algemene wet bestuursrecht, Article 6:7).

A letter from the inspectorate causes unrest, and that is understandable. My first step is calm: what is asked, which period applies and which device is it about? Then I collect your documents and read along before you reply. I advise cooperating and repairing quickly, while keeping your right to object.

Note: The inspectorate asks for documents or announces an order or a fine. Your first reply can matter in any later objection or appeal. Have a Dutch product law lawyer read your draft before you answer.

What do you do when a Dutch distributor wants your test under its own brand?

Settle who is manufacturer on the label. A distributor selling under its own name or brand takes over the manufacturer’s obligations, unless an agreement names you as manufacturer on the label and makes you responsible (Article 16(1)(a)).

A request like this feels like an opportunity and an extra demand at once, so I take it in order: your documents and label, the role split, translation and repackaging, whether your test still falls under the transition, and only then the contract. I propose that you stay on the label as manufacturer, with written agreements on documentation, complaints and recall, as long as you hold the file and certification.

Providing the information in translation and changing the outer packaging, where needed to market the device in that Member State, are not a modification (Article 16(2)). The distributor needs a quality system for the translation (Article 16(3)) and must inform you and the authority of that Member State at least 28 days before making the device available (Article 16(4)).

What can you arrange yourself?

Most IVDR work is technical: classification, performance evaluation, technical documentation, quality system, UDI and Eudamed.

A regulatory adviser and your notified body do the technical work better than a lawyer. I say that openly, so you also know why I sometimes say: you do not need to hand this one to me. You can already appoint your person responsible for regulatory compliance (a micro or small enterprise may hire one externally, Article 15(2)) and have your instructions and label drafted in Dutch.

When do you need a Dutch lawyer for the IVDR?

When there is a counterparty or a risk in the chain: an own-brand request, a mandate, a letter from the inspectorate, a transition that depends on conditions and dates, or a damage claim.

SituationWhat is at stakeWhy a lawyerWhat we do first
Distributor wants its own brandobligations pass unless the contract keeps you on the label (Article 16(1)(a))role, cost and liability are contract questionsreview the draft, set roles
Appoint or end a representativejoint liability (Article 11)the mandate decides recourse between you and the representative, and whether documents stay available if a business stops (Wmh Article 5b)review the mandate
IVDD status and notified body datestransition only if Article 110(3c) is metdates behind and aheadcheck facts and evidence
Inspectorate or Minister writesorder or fine; six weeks to object or appealdeadlines; your reply must hold upread the file, secure the deadline
Supply interruption or damage claimnotice at least six months ahead (Article 10a); financial cover (Article 10(15)); see the EU product liability directivenotice must match your contractsalign notice and contracts

One agreement makes the most difference: who stands on the label as manufacturer, and with that carries the documentation, the notifications and the liability. I record that first, then how complaints and notifications pass through the chain, that documents stay available if a business stops, and indemnity and financial cover. I propose one chain with one framework: mandate, distribution agreement and indemnity aligned.

Example: a class C test and a Dutch distributor with its own brand

Settle the role split before the first order. Suppose, hypothetically, a manufacturer outside the EU sold a rapid test on its own declaration for years and now needs a notified body (assume class C). A Dutch distributor wants its own brand. Did the manufacturer apply and sign by 26 May 2026 and 26 September 2026? Is there a sound mandate? Who is manufacturer on the label?

How does MAAK Advocaten work on IVDR files?

MAAK Advocaten in Amsterdam can act for manufacturers, representatives, importers and distributors of in vitro diagnostics. We establish your role, review contracts and handle contact with the Dutch authority; we do not classify devices or prepare technical files.

In the first days we read your documents and secure deadlines. You deal directly with Martin Krüger.

Frequently asked questions

Does the IVDR apply to my product, and what must I arrange as a manufacturer or importer?

Yes, if you intend it for the in vitro examination of specimens from the human body to give information on a condition or a treatment (Article 2, point 2).

Do I need an authorised representative in the EU?

If you are not established in a Member State, yes: you need one authorised representative under a written mandate (Article 11(1) and (2)), jointly liable with you if you fail Article 10 (Article 11(5)).

What does the authority do in the Netherlands if my product is checked?

The Minister can impose an order subject to a penalty and an administrative fine (Wet medische hulpmiddelen, Articles 10, 12 and 14); the period for lodging an objection or appeal is six weeks (Awb, Article 6:7).

When should I call a Dutch lawyer about the IVDR?

When a distributor wants its own brand, your representative needs a mandate, the inspectorate writes, or your transition depends on conditions and dates.

Legal framework and sources

About the author. Martin Krüger is a Dutch product compliance attorney (advocaat) and partner at MAAK Advocaten. His areas include CE marking and product regulation. · Profile

Is a Dutch distributor or authority asking about your IVDR file? Put your situation to our product law partner in the Netherlands, or get in touch with Martin Krüger.

News & Insights
Which topics would you like to receive updates on?
Who are you?
Which industry or sector do you operate in? (Optional)